ECG
IEC
AAMI/AHA
Electrode placement
R (red)
RA (white)
Right deltoid or wrist.
L (yellow)
LA (black)
Left deltoid or wrist.
N (black)
RL (green)
Right thigh or ankle.
F (green)
LL (red)
Left thigh or ankle.
C/C1 (white)
V/V1 (brown)
4
th
intercostal space, right border of the
sternum.
C2 (white/yellow)
V2 (brown/yellow)
4
th
intercostal space, left border of the
sternum.
C3 (white/green)
V3 (brown/green)
Midway between C2/V2 and C4/V4.
C4 (white/brown)
V4 (brown/blue)
5
th
intercostal space, mid-clavicular line.
C5 (white/black)
V5 (brown/orange)
Left anterior axillary line at C4/V4 level.
C6 (white/purple)
V6 (brown/purple)
Mid-axillary line at C4/V4 and C5/V5 levels.
Checking the ECG measurement
1.
Check that the waveforms and parameter values are displayed when the cable is
connected to the patient.
About the ECG analog output signal
Maximum delay of the ECG analog output signal is 30 ms with the PDM and 15 ms
with E-modules. Pacemaker pulse indication is included when appropriate and it is
summed in to the ECG waveform.
ECG synchronization pulse delay from the R-wave peak is <35 ms for the PDM and
E-modules, with the exception of wide QRS (120 ms/0.5 mV).
For more information and detailed specifications, refer to the supplemental
information manual.
154
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...