Cardiac output
C.O. safety precautions
C.O. warnings
WARNING
All invasive procedures involve risks to the patient. Use aseptic
technique. Incorrect use of the catheter can lead to vessel
perforation. Follow catheter manufacturer's instructions.
WARNING
When using an electrosurgery unit, note that the
measurement cables do not incorporate means to protect
against burns in case of a defective ESU return electrode. To
avoid burns at the monitor measurement sites, ensure the
following:
●
Proper contact of the ESU return electrode to the patient.
●
ESU return electrode near the operating area.
●
Measurement electrodes, leadwires and probes far from
the surgical site and the ESU return electrode.
WARNING
The cardiac output measurement results may be erroneous
during electrosurgery.
WARNING
During atrial fibrillation the C.O./CCO readings may be
erroneous.
C.O./CCO cautions
CAUTION
The C.O. time stamp indicates the time at which the C.O. value
was received by the monitor from the connected device.
In cases when the C.O. device is disconnected and then
reconnected to the Unity Network Interface Device (ID), the
time stamp may not indicate the actual time of the reading.
C.O. measurement limitations
●
E-modules used for this measurement are not suitable for use with neonatal
patients.
C.O. points to note
●
Cardiac output measurements from the PDM and TRAM require a C.O. license.
2062971-001
CARESCAPE Modular Monitors
261
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...