Messages
Message
Location
Explanation
What to do
●
Call service: Error x
●
where x = 0 - 99
●
param.
0 = RAM failure
1 = ROM checksum failure
2 = +15V failure
3 = -15V failure
6 = ADC error
7 = Watchdog time too short
8 = Watchdog time too long
9 = Watchdog activated
10 = EEPROM checksum error
11 = Autozero range exceeded
12 = Communication watchdog
activated
13 = Not in use
14 = Too early autocycle start
●
Contact authorized service
personnel.
●
Check NIBP
●
al. area
Systolic and/or diastolic results
missing.
●
Check the patient status.
●
Check NIBP cuff and hoses.
●
Repeat the measurement.
●
Control measurement
●
al. area
Pressure alarm limit exceeded.
●
Allow measurement to
complete.
●
Check the patient status.
●
Cuff occlusion
●
param.
Occlusion during measurement
or overpressured cuff.
●
Check the cuff.
●
Cuff loose
●
param.
Loose cuff or cuff hose.
●
Check the cuff and cuff hose.
●
Cuff overpressure
●
param.
NIBP cuff is squeezed during
measurement.
●
Check NIBP cuff and hoses.
●
Repeat the measurement.
●
Incorrect infl. limits
●
param.
Adult or child cuff is used, but the
selected infant mode restricts
the inflation pressure too low to
be able to measure the blood
pressure.
●
Adjust NIBP setup.
●
Long meas. time
●
param.
The measurement time is long.
The triggering values vary
according to the module and
inflation limits in use:
●
PSM: >2 min for adult/ child,
75 s to 80 s for infant
●
PDM: >2 min for adult/ child,
85 s for infant
●
TRAM: <3 min for adult/ child,
90 s for infant
●
Check the patient status.
●
Check the cuff and hose
connections.
●
Restart the measurement.
●
If the problem persists,
contact authorized service
personnel.
●
NIBP air leakage
/
NIBP manual
●
al. area
Loose cuff or cuff hose.
●
Check the cuff and cuff hose.
2062971-001
CARESCAPE Modular Monitors
473
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...