Bispectral index
BIS measurement setup
BIS equipment to patient connection
1.
Module with BIS measurement capability, E-BIS
2.
Digital Signal Processing Unit, BISx
3.
Patient Interface Cable, PIC Plus
4.
BIS sensor
BIS module keys
There are two keys on the module:
Opens or closes the BIS menu on the screen.
Starts the manual sensor check.
Preparing the patient for BIS measurement
1.
Connect the digital signal processing unit (BISx) cable to the module.
2.
Connect the patient interface cable to the BISx.
3.
Secure the BISx to a convenient location, preferably close to the patient’s head.
4.
Clean the application site with alcohol and let dry.
5.
Place the BIS sensor on the patient; see sensor package for instructions.
6.
Connect the sensor to the patient interface cable.
7.
Observe the results of the automatic sensor check in the parameter window.
8.
The measurement starts automatically after the sensor has passed the check.
Checking the BIS measurement
1.
Check that the sensor/electrode passes the sensor/electrode check when you
are starting to monitor a new patient.
396
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...