Starting and ending monitoring
monitor alarm is set to escalating) and the following ECG settings of the telemetry
will be used:
●
HR, ST, and PVC alarm limits
●
PVC alarm status (on/off)
●
Pacemaker detection
●
Lead analysis
●
Va lead position
●
Primary lead
●
ECG waveform size
●
Arrhythmia detection level
●
ST analysis status (on/off)
When combination monitoring is started with a non-admitted telemetry patient, these
same settings from the monitor will be sent to telemetry. Additionally, the
Telemetry
Waveforms
printing location is sent to telemetry.
If the telemetry alarm priority is such that it is not supported by the monitor, it will be
mapped to the next higher priority available.
NOTE
Patient’s age affects the alarm limits and alarm priorities
and also the configuration of the ECG algorithm including
arrhythmia alarms in the combination monitoring mode.
For more information, see the ApexPro Telemetry System
Operator’s Manual.
When combination monitoring is started with a non-admitted telemetry patient, the
printout type selection will be sent to the telemetry transmitter.
CAUTION
Users should be aware that all waveforms may be delayed
up to 2 seconds in the combination monitoring mode. If the
delay needs to be avoided, the combination monitoring mode
should be discontinued and all waveforms should be acquired
via the hard-wired bedside monitor.
CAUTION
Users should be aware of a possible time discrepancy
between the waveforms from the telemetry device and the
waveforms from a device hard-wired to the monitor. Users
should not consider these waveforms to be synchronous. If
absolute synchronicity is desired, the combination monitoring
mode should be discontinued and the ECG waveforms should
be acquired via the hard-wired bedside monitor.
Entering patient data
Entering patient data with the monitor
1.
Select the patient information area on the screen, or select
Data & Pages
>
Admit/Discharge
or
Start / Reset Case
.
2.
Select the
Patient
tab.
118
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...