Temperature
Temperature channel
Measurement source, B850
Measurement source, B650
or B450
T4
E-PT module
E-PT module
Tblood
PDM (Masimo), PDM (Nellcor),
TRAM 451, TRAM 451N5,
TRAM 451M, TRAM 851, TRAM
851N5, TRAM 851M, E-COP,
or E-COPSv module or Unity
ID.
PDM (Masimo), PDM (Nellcor),
E-COP, or E-COPSv module or
Unity ID.
Starting the temperature measurement
Connect the temperature probe to start the measurement. If the parameter window
displays
Off
in the value field:
1.
Select the temperature parameter window.
2.
Confirm that
Measurement
>
On
is selected.
Changing the temperature site label
1.
Select the temperature parameter window.
2.
Choose a site label from the
Label
list.
NOTE
In addition to the default channel labels, you can configure up
to six user-defined channel labels in the care unit settings. See
the CARESCAPE Modular Monitors Supplemental Information
Manual for details.
Displaying the delta value between two temperature
channels
NOTE
This selection is available when two temperatures are
displayed in the same temperature parameter window.
1.
Select the temperature parameter window.
2.
Select
Show Tx-Ty
(e.g.,
T2-T1
).
Setting temperature alarms
1.
Select the temperature parameter window.
2.
Select
Alarms
.
3.
Choose a temperature channel or temperature delta.
NOTE
If the feature is not active, the alarm limits are greyed out.
Select
Alarm On
to set the alarms.
4.
Set the alarm limits.
NOTE
The high limit alarm setting for delta values is also adjustable.
NOTE
If the setting of an alarm limit has been disabled during
configuration, the setting is marked with a lock symbol.
258
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...