Cardiac output
8.
Observe the washout curve displayed on the screen. The message
C.O. Complete
displays after the C.O. determination has been made, followed shortly by the
message
Please wait
.
The curve remains on the screen.
9.
Allow 1 to 1.5 minutes between injections to stabilize the catheter baseline
temperature.
10. To perform another C.O. measurement, wait for the
Press Start C.O.
message to
display, then select
Start C.O.
C.O. trial measurements
A real-time washout curve and numeric value are displayed with each cardiac output
trial. Up to six measurements are retained. The program automatically averages each
C.O. injection. When saved, the averaged value is entered into the hemodynamic
calculations. The last saved average C.O. value is displayed in the parameter window
with a timestamp.
Printing a report of C.O. trials
The C.O. report must be initiated before confirming the C.O. measurements.
1.
Select the cardiac output parameter window.
2.
Select the
Measurement
tab.
3.
Select
.
The selection is available only if you have not confirmed the C.O. measurements.
Editing the C.O. average
1.
Select the cardiac output parameter window.
2.
Select the
Measurement
tab.
3.
Select
Edit Average
.
4.
Check the selection boxes for those trials you wish to include in the C.O. average.
If you do not wish to include a trial, ensure its selection box is not checked.
5.
Select
Confirm C.O.
to store the calculated C.O. average and display it in the
cardiac output parameter window.
If you wish to print a C.O. report, you must start printing before confirming the C.O.
Canceling a C.O. measurement
When a C.O. measurement has just completed, you can remove this C.O. measurement
trial without entering the
Edit Average
window.
1.
In the
Measurement
tab, select
Cancel/Reject Injection
.
E-modules: In addition to removing the previous measurement, you can also
cancel an in-process measurement.
C.O. catheter selections
You can select a cardiac output catheter from a list of default catheters and
preconfigured catheters, or enter a catheter for temporary use. Pre-configured
266
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...