Impedance respiration
Problem
Solution
●
Other electrical devices may interfere with the
measurement.
Why does the waveform have a combination of
shallow and deep breaths, but the monitor is not
detecting the shallow breaths?
If the detection sensitivity threshold is set too high,
shallow breaths will not be detected, as shown in the
following example of incorrect detection (1 = breath).
●
Decrease the detection sensitivity percentage until
the markers correctly identify each inspiration and
expiration or set to
AUTO
(PSM). If the detection
mode is
AUTO
, the grid lines represent the
minimum limits. The limits in use may be a larger
range. The following is an example of correct
detection.
Respiration detection is not dependent on the size of
the waveform. Size is for visual purposes only.
Why is the monitor detecting cardiac artifact as
breaths?
The breath detection threshold is too low (1 = breath,
2 = artifact). The following is an example of incorrect
detection.
●
Increase the detection sensitivity percentage until
the markers correctly identify each inspiration and
expiration. The following is an example of correct
detection.
200
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...