Cardiac output
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select the radio button for
Automatic
measurement type.
4.
Verify that the catheter settings are correct.
5.
Select the
Measurement
tab.
6.
Complete the following:
a. PDM and TRAM: Get ready to inject the injectate solution when the message
Inject When Ready
appears.
b. E-modules: When the message
Press Start C.O. Serial
appears, select
Start
C.O. Serial
.
7.
Inject the injectate solution smoothly within 4 to 5 seconds.
8.
The message
Measuring
displays, followed by the message
Please wait
until
the calculation is completed.
9.
Observe the washout curve displayed on the screen. The message
C.O. Complete
displays after the C.O. determination has been made, followed shortly by the
message
Please wait
.
The curve disappears from the screen when the next measurement cycle can
start.
10. To take another C.O. measurement, wait for this message to display before
injecting the injectate:
●
PDM and TRAM:
Inject When Ready
●
E-modules:
Inject now!
Taking a manual C.O. measurement
Measuring C.O. using the manual mode allows you to determine when to begin the
injection procedure. This mode may be preferred for patients with extreme blood
temperature fluctuations, or when the automatic mode is unable to establish a stable
baseline.
Holding the injectate syringe by the plunger, not the barrel, improves measurement
accuracy.
1.
Select the cardiac output parameter window.
2.
Select the
Setup
tab.
3.
Select the radio button for
Manual
measurement type.
4.
Select the
Measurement
tab and verify that the catheter settings are correct.
5.
Select
Start C.O.
With E-modules, you can also use the
Start C.O.
module key.
6.
When the
Inject now!
message appears, inject the injectate solution smoothly
within 4 to 5 seconds.
7.
The message
Measuring
displays, followed by the message
Please wait
until
the calculation is completed.
2062971-001
CARESCAPE Modular Monitors
265
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...