Airway gases
Using the CARESCAPE respiratory modules and
compact airway modules for O₂ measurement
Selecting the O
2
scale
If the difference between FiO
2
and EtO
2
is more than 6%, change the O
2
scale.
1.
Select a gas related parameter window.
2.
Select the
O2
tab >
Setup
.
3.
Select an option from the
Scale
list.
Selecting the O
2
sweep speed
This selection affects the waveform.
1.
Select a gas related parameter window.
2.
Select the
O2
tab >
Setup
.
3.
Select an option from the
O2 Sweep Speed
list. The options are
0.625 mm/s
,
6.25 mm/s
,
12.5 mm/s
, and
50 mm/s
.
The smaller the value, the slower the sweep speed.
Setting O
2
alarms
1.
Select a gas related parameter window.
2.
Select the
O2
tab >
Alarms
.
3.
Check that the required alarm (
EtO2
or
FiO2
) is on, and set its high and/or low
limit values.
Using the CARESCAPE respiratory modules
and compact airway modules for AA and N₂O
measurement
Selecting the agent scale
Every anesthetic agent has its own default scale that the monitor uses when detecting
the agent. You can change the scale of an agent if the amount used is higher than
the default scale. Default scales are given in the CARESCAPE Modular Monitors
Supplemental Information Manual.
1.
Select a gas related parameter window.
2.
Select the
Agent/N2O
tab >
Setup
.
3.
Select an option from the
Agent Scale
list.
Selecting the agent sweep speed
This selection affects the waveform.
294
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...