Invasive pressures
Invasive pressure calibration with E-modules
The invasive pressure calibration with E-modules requires specific tools and setup. For
detailed instructions, see the Module Frames and Modules Technical Manual.
Invasive pressure practicalities
Invasive pressure parameters
The measured invasive pressure parameters are systolic, diastolic, and mean. Pulse
rate can be monitored with any arterial site. PCWP can also be measured for a PA
site. CPP is a calculated value that requires a valid ICP value and a valid arterial site
value. In addition, also SPV and PPV can be measured.
With the B850 you can monitor up to eight pressures, with the B650 you can monitor
up to seven pressures, and with the B450 you can monitor up to six pressures.
The following table lists the available site names and displayed values:
Site
Site name
PDM and TRAM,
displayed values
E-modules, displayed
values
General site name for the
specific invasive pressure
channels 1 to 8
P1
to
P8
Mean
Systolic, diastolic, mean
Arterial
Art
Pulse rate, systolic,
diastolic, mean
Pulse rate, systolic,
diastolic, mean
Arterial blood
ABP
Pulse rate, systolic,
diastolic, mean
Pulse rate, systolic,
diastolic, mean
Central venous
CVP
Mean
Systolic, diastolic, mean
Femoral arterial
Fem
Pulse rate, systolic,
diastolic, mean
Pulse rate, systolic,
diastolic, mean
Femoral venous
FemV
Mean
Systolic, diastolic, mean
Intracranial
ICP
Mean
Systolic, diastolic, mean
Left atrial
LAP
Mean
Systolic, diastolic, mean
Pulmonary artery
PA
Systolic, diastolic, mean
Systolic, diastolic, mean
Right atrial
RAP
Mean
Systolic, diastolic, mean
Right ventricular
RVP
Mean
Systolic, diastolic, mean
Umbilical artery
catheter*
UAC
Pulse rate, systolic,
diastolic, mean
Not supported.
Umbilical venous
catheter*
UVC
Mean
Not supported.
* PDM and TRAM with NICU software package only.
Intra-aortic balloon pump
NOTE
PDM, TRAM, and Tram-Rac only. Not available in the NICU
software package.
2062971-001
CARESCAPE Modular Monitors
249
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...