Pulse oximetry
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
tab.
3.
Select
Show Sat. Seconds
.
Setting the Saturation Seconds threshold
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
tab.
3.
Set the threshold with the
Saturation Seconds
arrows.
Setting the SpO
2
alarms and alarm limits
You can set the alarms and alarm limits for primary and secondary SpO
2
measurements separately.
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
or
SpO2(2)
tab.
3.
Select the
Alarms
tab.
4.
Set the alarm limits for the
SpO2
,
HR
, or
PR(SpO2)
.
If a feature is not active, alarm limits are greyed out. Select
Alarm On
to set the
alarm limits.
HR
appears when the
HR Alarms
is set to
Single
.
PR(SpO2)
appears
when the
HR Alarms
is set to
Multiple
. The
HR
and
PR(SpO2)
settings are not
available for the secondary SpO
2
measurement.
5.
Set the alarm limits.
Deactivating the
SpO2 probe off
alarm
This feature is meant to be used when ending SpO
2
monitoring. It should not be used
during active SpO
2
monitoring. This setting can be enabled during configuration. If it
has been enabled, there will be a button on the
SpO2
and
SpO2(2)
tabs that allows you
to deactivate the alarm:
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
or
SpO2(2)
tab.
3.
Select
Deactivate SpO2 Probe Off
.
When the alarm is deactivated, there will be no audible or visual
SpO2 probe off
alarm
indications. The alarm is automatically reactivated if SpO
2
vitals signs are detected
and alarm condition is met again.
WARNING
With deactivated
SpO2 probe off
alarm, keep the patient
under close surveillance.
SpO
2
alarm priorities
You can select priorities for the
SpO2 high
,
SpO2 low
, and
SpO2 probe off
alarms
through
Alarm Setup
>
Alarm Priorities
>
Other Parameters
. The choices are
Low
,
Medium
,
High
, and
Escalating
.
Stopping the SpO
2
measurement
1.
Remove the SpO
2
sensor from the patient.
2062971-001
CARESCAPE Modular Monitors
213
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...