Cardiac output
4.
Make necessary changes by selecting the
Hemo
,
Oxy
, or
Vent
tab and then
Edit
Input
.
5.
You can return to the
Cardiac Output
menu by selecting
C.O.
or
Previous Menu
.
Adjusting the wedge from the cardiac output menu
The
Wedge
selection is available only when there is a confirmed C.O. measurement
and an invasive pressure channel has been labeled as
PA
.
1.
Select the cardiac output parameter window.
2.
Select the
Measurement
tab.
3.
Select
Wedge
.
4.
Adjust the wedge settings as required.
5.
You can return to the cardiac output menu by selecting
Previous Menu
.
Stopping the cardiac output measurement
1.
Remove the catheter from the patient.
2.
Disconnect the probe patient cables.
C.O. practicalities
Cardiac output washout curve
The washout curve, which displays after a C.O. injection, shows the drop in blood
temperature as the injectate mixes with the blood. The peak of the curve indicates the
maximum difference in the patient’s baseline blood temperature and the temperature
of the injectate solution. As the mixture passes through the catheter and then out the
pulmonary artery, the temperature difference decreases as indicated by the curve
returning to the baseline. A spike is displayed at the onset of the curve, again at
70% of the maximum temperature difference, and again at 35% of the maximum
temperature difference. Spikes are also visible during an ongoing C.O. measurement.
1.
Onset spike
2.
70% spike
3.
35% spike
Cardiac output is inversely proportional to the area under the thermodilution curve.
Cardiac output varies with body size. To more accurately assess cardiac performance
for individual patients, the cardiac index is often used.
How to improve the C.O. accuracy
The following influencing factors can influence the cardiac output accuracy:
2062971-001
CARESCAPE Modular Monitors
269
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...