2.
Select
Start with
>
Current patient
.
3.
Select
Continue
.
Restarting the NMT measurement in OR after induction
When you move the patient with the module to the OR and want to continue the
measurement with the already found and determined current and reference values,
use the restart function. Connect the module to the monitor and press the
Stop
Continue
module key, or:
1.
Select the NMT parameter window >
Setup
.
2.
Select
Start with
>
Recall reference
.
3.
Select
Restart
.
Stopping the NMT measurement
Press the
Stop Continue
module key, or:
1.
Select the NMT parameter window >
Setup
.
2.
Select
Stop
.
NMT alternative uses
Local nerve and plexus localization
Single stimulation pulses may be helpful for determining the correct needle tip position
from a local nerve in plexus procedures. The NMT module delivers single stimulation
pulses at a selected rate until it is manually stopped. Muscle contractions from the
stimulated innervating motoric nerve(s) may be observed. Note that you need specific
accessories for this measurement.
2062971-001
CARESCAPE Modular Monitors
367
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...