30
Viewing other monitored patients
About viewing other monitored patients
When the monitor is on the network, you can open a bed-to-bed view of other remote
patient beds that are on the same network. You can choose to view a remote patient
bed under an alarm condition, or simply view any available bed on your network.
The numeric values, up to six waveforms, alarms, and location information are
displayed inside a separate bed-to-bed window. The bed-to-bed window is located on
the left side of the display screen.
Function
Network features
View on alarm notification
CARESCAPE Network and S/5 Network: Monitor alarms for
up to 40 beds.
View remote beds
CARESCAPE Network: View one bed from up to 102
3
beds.
S/5 Network: View one bed from up to 128 beds.
NOTE
When using WLAN connection, the maximum number of
remote beds is five (B650) or ten (B450).
Some settings related to remote alarm configuration are given through
Care Unit
Settings
>
Alarms
>
Remote Alarms
, and they are password protected:
●
Enabling or disabling audio pausing remotely for another monitor (
Allow Audio
Pause: For Remote Bed
). Choices are
Yes
and
No
. CARESCAPE Network only.
●
Selecting which remote locations are allowed to pause audio alarms on this
monitor (
Allow Audio Pause: From Remote Location
). Choices are
Not Allowed
,
Central
, or
Central and Remote Beds
. CARESCAPE Network only.
●
Selecting alarm priorities that can be paused remotely (
Allow Remote Pausing of
).
Choices are:
Low Alarms
,
Low & Medium
, or
All Alarms
.
●
Display of the remote patient name (
Show Remote Patient Name
check box).
●
Use of the alarm light for a remote alarm (
Remote Alarm Light
). Choices are
On
or
Off
.
●
Selecting the remote alarm tone (
Remote Alarm Tone
). Choices are
Off
,
Single
,
Repeat
, or
Local
.
●
Enabling or disabling the restoring of the remote bed selections after discharge
(select or deselect the
Restore after Discharge
check box for
Remote Bed
Selections
).
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
2062971-001
CARESCAPE Modular Monitors
443
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...