Invasive pressures
Module
Key
Functionality
E-modules
Zero P1
to
P8
Zeros the reference for each pressure transducer
individually.
PDM
Zeros the reference for all pressure transducers
connected to the PDM.
Connecting the invasive pressure transducer and cable
1.
Prepare the transducer kit according to the manufacturer’s instructions.
2.
Connect the pressure transducer to the transducer cable.
3.
Remove entrapped air from within the transducer setup.
4.
Connect the transducer cable to the acquisition module’s invasive pressure
connector.
5.
Connect the transducer to the patient line.
Checking the invasive pressure measurement
1.
Check that the monitor recognizes cable connections (activates the display) for all
the pressure channels used and the pressure values and appropriate waveforms
are displayed.
2.
Make sure that all the transducers are zeroed correctly.
Invasive pressure measurement on the monitor screen
The invasive pressure channel labels are as follows:
Label
Description
Art
Arterial pressure
ABP
Arterial blood pressure
Fem
Femoral arterial pressure
FemV
Femoral venous pressure
PA
Pulmonary arterial pressure
CVP
Central venous pressure
LAP
Left atrial pressure
RAP
Right atrial pressure
ICP
Intracranial pressure
RVP
Right ventricular pressure
UAC
Umibilical arterial pressure
UVC
Umbilical venous pressure
P1
to
P8
Non-specific pressure channel labels
NOTE
UAC
and
UVC
invasive pressure channels are only available in the
NICU software package.
2062971-001
CARESCAPE Modular Monitors
237
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...