Messages
Messages related to EEG and AEP measurements
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
AEP stimulation on
●
param.
The measurement is being
performed and clicks are being
delivered to the headphones.
●
Wait until the stimulation is
completed.
●
Artifacts
●
param.
Electrodes are poorly connected,
or electrical interference is
coupling to electrode cables.
●
Check the electrodes and
their impedances.
●
Remove the sources of noise,
if necessary.
●
Big contact difference
●
param.
The electrode impedance check
fails.
●
Check the contact quality
from the active and
baseline (+ and -) electrodes
establishment to ensure
measurement from the same
noise signal.
●
Perform a new electrode
impedance check.
●
Check GND electrode
●
param.
The ground electrode
impedance is over 5 kOhm.
●
Check the contact quality
from the ground electrode
establishment.
●
If necessary, change the
electrode.
●
Perform a new electrode
impedance check.
●
Checking electrodes
●
param.
Electrode impedance is being
measured and EEG analysis is
stopped for a few seconds.
●
Wait until the electrode
impedance check is
completed.
●
EEG leads off
●
Leads off
●
al. area
●
param.
Leads are not connected.
●
Reconnect the disconnected
headbox leads.
●
EEG measurement off
●
param.
The measurement has been
stopped.
●
Restart the measurement if
required.
●
EEG measurement
removed
●
al. area
Acquisition module has been
removed.
●
Connect the module if
you want to restart the
measurement.
●
Electrodes OK
●
param.
Electrode impedance is below 5
kOhm.
●
You can continue the
measurement.
500
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...