Trends
2.
Select
Numeric
from the
View
list.
●
To see other parameters, select their tabs in the
Numeric
trend view.
●
To see more numeric trend data, use the cursor to scroll the data in horizontal
direction.
Changing the time interval of numeric trends
Numeric trends display values according to the selected time interval. Numeric
trends are updated with averaged measurement data once a minute independent of
the selected time scale.
1.
Select
Trends
.
2.
Select
Numeric
from the
View
list.
3.
Select a value from the
Time Interval
list.
For example, a 5 minute interval will show data for every 5 minutes, and a 30 minute
interval will show data for every 30 minutes. The data is displayed in columns on the
screen. NIBP, PCWP, cardiac output, NMT, and manual SPV measurements will always
add one column independent of the
Time Interval
setting.
Printing numeric trends
1.
Select
Trends
.
2.
Select
Numeric
from the
View
list.
3.
Select
Print Page
(Recorder).
4.
You can stop printing by selecting
Stop Printing
.
Depending on what has been configured, either the data currently on screen or all
data will be printed. This configuration is done through
Monitor Setup
>
Default Setup
>
Care Unit Settings
>
Printer
>
Numeric Trends Printing
and it is password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Invasive pressure trends
The following invasive pressure trend data is collected:
●
TRAM or PDM and Dual BP, BP, BP/Dual temperature single-parameter Tram-Rac
modules:
■
Systolic, diastolic, and mean pressure data for
Art
,
ABP
,
Fem
,
PA
, and
UAC
.
■
Mean pressure data for
FemV
,
CVP
,
LAP
,
RAP
,
ICP
,
P1
to
P8
,
RVP
, and
UVC
.
●
E-modules:
■
Systolic, diastolic, and mean pressure for all labels except
UAC
and
UVC
.
The invasive pressure trends will only be stored for those channels that have been
zeroed.
Heart rate (HR) trends
Only the measured values from the primary heart rate or pulse rate source are trended
to the HR graphic and numeric trends. Pulse rate sources that can be selected are the
420
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...