Invasive pressures
Problem
Solution
The measurement values in the SPV and PPV
parameter window are not displayed.
●
Ensure that there are no arrhythmias present.
●
Check that the correct arterial site has been
selected as the SPV source.
●
If you are using the SPV measurement in the NICU
software package, only manual measurement is
shown.
●
The algorithm requires at least three pulse beats
per each respiration cycle and when this is not
true, the values are not shown. Try using manual
measurement.
Why are displayed pressure values different than
expected?
●
Check the patient. Values could be valid, the
patient could be lying on the tubing, or the tubing
could be kinked.
●
Check tubing for bubbles.
●
Remove excess tubing.
●
Check phlebostatic axis placement of transducer.
●
Rezero pressure.
●
If patient is on IABP, verify that the monitor’s IABP
program is turned on. If necessary, turn it on.
Why are the arterial, non-invasive (oscillometric),
and auscultated blood pressure readings indicating
different values?
The three measurement methods use different
technologies. Auscultation and oscillometric are
both indirect methods of measuring blood pressure.
In auscultation, changes in arterial sounds during
cuff deflation are related to systolic and diastolic
pressure. With oscillometric measurement, changes
in measured pressure oscillations during cuff
deflation are related to systolic, mean and diastolic
pressures. Changes in the vascular tone of the
arterial system can cause these two indirect methods
to differ from one another and from direct arterial
pressure measurements.
Invasive arterial blood pressure is a direct method
of measuring blood pressure. Differences between
direct and indirect blood pressure measurements are
expected. These differences occur because direct
methods measure pressure and indirect methods
measure flow. In addition, differences occur because
the measurement location is not the same (e.g.,
brachial artery for NIBP vs. radial artery for invasive
arterial pressure monitoring).
2062971-001
CARESCAPE Modular Monitors
253
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...