ECG
3.
Select the
Location ID
field and enter the ID with the on-screen numeric keypad.
You can enter any number from 0 to 599.
Setting automatic 12 lead ECG analysis measurements
1.
Select the HR parameter window.
2.
Select
12 Lead Analysis
.
3.
Select
Settings
.
4.
Select a time interval from the
Auto Interval
list.
5.
Select
Cancel
or
Confirm
.
Setting the 12 lead ECG analysis display format
This setting will change the 12 lead ECG waveform display format and the printed
12 lead ECG report waveform format.
1.
Select the HR parameter window.
2.
Select
12 Lead Analysis
.
3.
Select
Settings
.
4.
Select a format from the
Display Format
list:
●
4 x 2.5 - 1 Rhythm
●
4 x 2.5 - 3 Rhythms
●
12 Rhythms
●
Cabrera
5.
Select
Cancel
or
Confirm
.
Generating a 12 lead ECG analysis report during an ST
alarm condition
You can have a 12 lead ECG report automatically generated when an ST alarm
condition occurs. Automatically generated reports are viewable through
12 Lead
Analysis
>
Saved Reports
.
1.
Select the HR parameter window.
2.
Select
12 Lead Analysis
.
3.
Select
Settings
.
4.
Select
On
from the
12 Lead on ST Alarm
list.
5.
Select
Cancel
or
Confirm
.
Performing a 12 lead ECG analysis
1.
Select the HR parameter window.
2.
Select
12 Lead Analysis
.
3.
Select
12 Lead Now
.
All the waveforms in the
12 Lead Analysis
view freeze during the analysis except for
the ECG I waveform. Analysis takes less than one second to complete. At that time, the
2062971-001
CARESCAPE Modular Monitors
169
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...