Viewing other monitored patients
Automatic view of remote beds in alarm
Automatic viewing of remote alarms is a licensed feature.
You can set the monitor to automatically notify you with an alarm message or with a
bed-to-bed window when selected remote patient beds go into an alarm condition.
All automatically viewable alarming beds display in order from the highest to lowest
alarm priority and from the newest to oldest alarms.
You can configure how the monitor notifies you of a remote patient bed alarm
condition, and which remote patient alarm priority levels you want notification of. You
can do this for individual beds, or for all remote beds of a selected care unit at once.
Selecting the alarm notification type
1.
Select
Data & Pages
.
2.
Select
Other Patients
.
3.
Select the
Receive Alarms
tab.
4.
Select a care unit from the
Unit
list.
5.
Select a patient bed from the displayed list.
6.
Select the type of alarm notification desired from the
Alarm Notification
list:
●
Off
: Remote alarm notification is turned off.
NOTE
All new monitors appear on the list automatically with
the notification setting
Off
(default).
●
Message
: Remote alarm messages display in the alarm area. At any time,
you can select a remote alarm message to open a bed-to-bed window and
view the remote patient’s data.
●
Auto View
: Bed-to-bed window opens immediately if no other procedure or
setup window is currently open. Otherwise, a remote alarm message displays
in the alarm message area. To open the bed-to-bed window and view the
remote patient’s data, close the currently open menu, or select the remote
alarm message.
●
Auto View Always
: Immediately closes any open procedure or setup windows
and opens a bed-to-bed window with the remote patient’s data.
Selecting the notifying alarm priority level
1.
Select
Data & Pages
.
2.
Select
Other Patients
.
3.
Select the
Receive Alarms
tab.
4.
Select a care unit from the
Unit
list.
5.
Select a patient bed from the displayed list.
444
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...