Monitoring basics
1.
Tram-Rac housing
2.
Release levers
Connecting and removing modules: B650
WARNING
When connecting PDM, the loaded IP labels may affect the
channel labeling of other already connected channels, and
consequently also the alarm limits.
Removing a PSM or PDM from the B650
1.
Pull the release tab out and slide the module out of the guides.
2.
Hold onto the module to make sure it does not drop when it comes out.
Connecting a PDM or PSM to the B650
1.
Connect a module by aligning it with the insertion guides on the pivoting module
frame.
2.
Push the module into the module frame until it clicks and stops.
NOTE
The PDM module requires additional time to power up when used
without the PDM battery. Do not interrupt the startup sequence
by unplugging the PDM module.
Connecting other E-modules than PSM to the B650
1.
With the module properly oriented (module release latch facing down), align the
insertion guide slot in the module with the insertion guide in the module frame.
2.
Push the module into the module frame until it clicks.
96
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...