Airway gases
WARNING
Do not use a CO
2
module at the same time as a compact
airway module or a CARESCAPE respiratory module.
WARNING
When using the CARESCAPE respiratory modules with volume
controlled ventilation at low tidal volumes, the specified gas
withdrawal rate may significantly reduce the amount of gas
delivered to the patient.
WARNING
CARESCAPE respiratory modules: Make sure to compensate
for the possible reduction of tidal volume caused by the 120
ml/min gas sample flow.
WARNING
Compact airway modules: Do not use these modules on
patients that cannot tolerate the removal of 200 ml/min from
their total minute ventilation.
WARNING
E-miniC modules: Do not use this module on patients that
cannot tolerate the removal of 150 ml/min from their total
minute ventilation.
WARNING
A failure in zeroing or calibrating airway gases may cause
inaccurate readings.
WARNING
Ensure that the compact airway modules and CARESCAPE
respiratory modules are in vertical position when used. Tilting
them may result in erroneous readings.
WARNING
Since calibration gas contains anesthetic agents, always
ensure sufficient ventilation of the room during calibration.
WARNING
Do not wash, disinfect or open the water trap cartridge.
Do not touch the water trap membrane. The hydrophobic
membrane is damaged if any cleaning is attempted, and this
may result in the contamination of the gas sensors.
WARNING
Compact airway modules: Never connect any tubing to the
reference gas inlet connector. The inlet must be open at all
times.
WARNING
E-miniC: O
2
, N
2
O and anesthetic agent gases may interfere
with EtCO
2
readings.
WARNING
To avoid the risk of patient cross-infection, do not return the
sampled gas to the breathing system.
282
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...