Mixed venous oxygen saturation (SvO₂)
SvO
2
points to note
●
Follow the catheter manufacturer's instructions for inserting the catheter.
●
For reliable saturation values, the signal strength indicator should be higher than
one asterisk.
●
Depending on the SvO
2
module used, not all SvO
2
measurements and settings
are available to view or change.
SvO₂ measurement setup
SvO
2
equipment to patient connection
1.
Module with SvO
2
measurement capability, E-COPSv
2.
Optical module
3.
Optical connector
4.
Swan-Ganz thermodilution catheter
Checking the SvO
2
measurement
1.
Check that the SvO
2
value is displayed when the catheter is inserted, the
measurement has been started and the
Warming up
message has disappeared.
2.
Check the position of the catheter regularly.
3.
Calibrate in vivo and update the Hb value at least every 24 hours.
SvO
2
measurement on screen
Signal quality indicator from the catheter (only present when the SvO
2
is measured
from an external device):
Indicator
Explanation
no asterisk
No signal
*
Poor signal
**
Fair signal
***
Good signal
276
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...