Safety
WARNING
— ELECTRIC SHOCK — Only use protected leadwires and
patient cables with this monitor. The use of unprotected
leadwires and patient cables creates the potential for making
an electrical connection to ground or to a high voltage power
source which can cause serious injury or death to the patient.
WARNING
For detailed instructions and information regarding supplies
and accessories, always refer to their own instructions for use.
Cables warnings
WARNING
— CABLES — Route all cables away from patient’s throat to
avoid possible strangulation.
WARNING
— SITE REQUIREMENTS — Do not route cables or tubing in a
way that they may present a stumbling hazard.
WARNING
— SAFETY GROUND — Remove power cord from the mains
source by grasping the plug. Do not pull on the cable.
Defibrillation warnings
WARNING
Do not touch the patient, table, bed, instruments, modules or
the monitor during defibrillation.
WARNING
— DEFIBRILLATOR PRECAUTIONS — Patient signal inputs
labeled with the CF and BF symbols with paddles are
protected against damage resulting from defibrillation
voltages. To ensure proper defibrillator protection, use only
the recommended cables and leadwires.
Electrical warnings
WARNING
— POWER SUPPLY — The device must be connected to a
properly installed power outlet with protective earth contacts
only. If the integrity of the protective earth conductor is in
doubt, disconnect the monitor from the power line (and use it
with the battery option (B650) if available). If the installation
does not provide for a protective earth conductor, disconnect
the monitor from the power line. All devices of a system must
be connected to the same power supply circuit. Devices which
are not connected to the same circuit must be electrically
isolated when operated.
WARNING
— EXCESSIVE LEAKAGE CURRENT — Do not use a multiple
socket outlet or extension cord.
WARNING
— EXCESSIVE TOUCH CURRENT — To avoid excessive patient
leakage current, do not simultaneously touch the patient and
the electrical connectors located at the rear panel of the CPU
unit or monitor, or within the module housing or frames.
WARNING
— INTERFACING OTHER EQUIPMENT — Connect only items
that are specified as part of the system and as compatible.
For more information, see the CARESCAPE Modular Monitors
Supplemental Information Manual.
44
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...