CO2 with CAPNOSTAT Mainstream, CapnoFlex LF, and Dual CO2 modules
9.
Check the monitor screen. If the message
Calibrate sensor
appears, calibrate the
sensor. If the message
Check/calibr. adapter
appears, calibrate the adapter.
10. Block the open end of the sample line or nasal cannula. Within 90 seconds the
pump will shut off and the message
Sample line blocked
should appear on the
monitor display to ensure that the monitor recognizes blockages.
11. To turn the pump back on and return to normal operation, turn the pump off and
then back on again in the CO
2
setup menu. The pump will turn back on and the
module will return to normal operation.
12. Connect the sample line to the patient. Ensure that the tubing is not hanging
lower than the patient.
NOTE
Delivery of bronchodilators and/or mucolytics via aerosol and
meter dose inhalers can be corrosive and cause premature
blockage of the water trap. Discontinue sidestream gas
analysis prior to treatment by switching the pump off and
removing the sample line from the breathing circuit.
Calibrating the Dual CO2 sensor
Sensor calibration should be performed whenever the message
Calibrate sensor
appears.
1.
Place the sensor away from all sources of CO
2
, including the patient’s exhaled
breath, your exhaled breath, and ventilator exhaust valves.
2.
With the cable connected to the module, select a gas related parameter window >
Calibration
.
3.
Place the sensor on the cell marked
0
and select
Start Calibrating
.
4.
After the message
Zeroed
appears, remove the sensor from the
0
cell and attach
it to the cell marked
REF
.
The measured EtCO
2
value appears in the
Connect sensor to REF cell
window. If
the value stays between 36 and 40 mmHg for three seconds during the 20-second
calibration, the calibration is successful and the message
Calibrated
appears. If the
calibration fails, the message
Recalibrate
appears instead. In this case, perform
a new calibration.
Calibrating the Dual CO2 adapter
Adapter calibration should be performed whenever the message
Check/calibr.
adapter
appears.
1.
Place the sensor and adapter away from all sources of CO
2
, including the patient’s
exhaled breath, your exhaled breath, and ventilator exhaust valves.
2.
With the adapter connected to the sensor and the sensor connected to the
monitor (and, when used, all sidestream tubing attached and the pump turned
on), select a gas related parameter window.
3.
Select the
Calibration
tab.
4.
Select
Start Calibrating
.
The message
Calibrated
appears on the screen after a successful calibration. If the
calibration fails, the message
Calibration error
appears instead. In this case, perform
a new calibration.
314
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...