Messages
Message
Location
Explanation
What to do
●
Replace sensor
●
param.
The sensor date has expired, the
sensor has been used too many
times, or the validity time for the
sensor cannot be determined.
●
Replace the sensor.
●
Sensor check failed
●
param.,
wavef.
The sensor has not passed the
impedance check.
●
Check sensor placement and
attachment.
●
Press each electrode in the
sensor.
●
Replace the sensor.
●
Testing DSC
●
param.,
wavef.
The BISx test has been activated.
●
Wait until the check is
completed.
Messages related to TC measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Calibrating
●
param.
Calibration is in progress.
●
No action required.
●
Calibration error
●
param.
Unsuccessful calibration.
●
Authorized service personnel
should repeat the calibration
procedure.
●
Connecting
●
param.
Connection to the interfaced
device is being established.
●
No action required.
●
Ext device error
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
No ext device
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Not calibrated
●
param.
Measurement needs to be
calibrated.
●
Perform calibration.
●
Ready
●
param.
Measurement is ready.
●
No action required.
2062971-001
CARESCAPE Modular Monitors
503
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...