3
System introduction
Monitor and module compatibility limitations
TRAM and Tram-Rac modules with the B850 only.
System safety precautions
System warnings
WARNING
Never install equipment above the patient.
WARNING
Operation of the monitor outside the specified performance
range may cause inaccurate results.
WARNING
— EXCESSIVE LEAKAGE CURRENT — A secondary display or
printer that is a non-medical grade device and is used within
the patient environment, must always be powered from
an additional transformer providing at least basic isolation
(isolating or separating transformer). Using without an
isolating transformer could result in unacceptable enclosure
leakage currents.
WARNING
— EXCESSIVE LEAKAGE CURRENT — Laser printers are UL
60950/IEC 60950 certified equipment, which may not meet the
leakage current requirements of patient care equipment. This
equipment must not be located in the patient environment
unless the medical system standard IEC 60601-1-1 is followed.
Do not connect a laser printer to a multiple socket outlet
supplying patient care equipment. The use of multiple socket
outlet for a system will result in an enclosure leakage current
equal to the sum of all the individual earth leakage currents
of the system if there is an interruption of the multiple socket
outlet protective earth conductor. Consult your local service
representative before installing a laser printer.
WARNING
— PHYSICAL INJURY — Take care when mounting devices to
an IV pole. If a device is mounted too high the IV pole may
become unbalanced and tip over.
2062971-001
CARESCAPE Modular Monitors
53
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...