Impedance respiration
Respiration measurement setup
Respiration equipment to patient connection
1.
Module with ECG measurement capability
2.
AAMI/AHA or IEC Multi-Link 3/5-lead, 6-lead, or 12SL ECG cable
3.
AAMI/AHA or IEC 3-leadwire, 5-leadwire, or 6-leadwire set
4.
AAMI/AHA or IEC precordial leads leadwire set
Preparing the patient’s respiration electrode sites
Excessive body hair or skin oil reduces electrode contact with the skin and decreases
the quality of electrode signal.
When preparing the electrode sites, avoid obvious layers of fat and major muscles.
1.
Shave any hair from the electrode site.
2.
Gently rub the surface of the skin to increase capillary blood flow.
3.
Clean the skin with alcohol or a mild soap and water solution to remove skin oil
and dead or abraded skin cells.
4.
Dry the skin completely before applying the electrodes.
Respiration lead and breath detection
Respiration leads identify the ECG leads used for respiration measurement. Each
respiration lead is suited for specific breath detection conditions:
2062971-001
CARESCAPE Modular Monitors
193
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...