Patient Spirometry
●
x = Time
●
y =
1. Inspiration
2. Expiration
Compliance
The measurement of pressure and volume allows for the calculation of the lungs’
dynamic compliance - in other words, how well an additional volume enters the lungs
when pressure is exerted on the airway. Compliance describes the extensibility
of the lung-thoracic system but is dominated by alveolar (that is, purely peripheral)
properties.
Typical compliance for adults varies between 35 ml/cmH
2
O and 60 ml/cmH
2
O, and
typical compliance for children is more than 15 ml/cmH
2
O.
Decrease in compliance may indicate too high a PEEP level, acute lung injury, or
increased intra-abdominal pressure (bleeding, ascites). Decreased compliance
indicates a worsening ventilatory status.
Typical compliance
●
x = Paw
●
y = Vol
Low compliance
●
x = Paw
●
y = Vol
Airway resistance (Raw)
Airway resistance expresses the relationship between the pressure difference across
the airway (between the mouth and the alveoli) and the rate at which gas is flowing
through the airway. An increasing airway resistance usually indicates an airway
obstruction and may lead to barotrauma. In critically ill patients, factors causing
increased airway resistance should be identified as quickly as possible to avoid lung
overdistension or decreased oxygen delivery.
NOTE
An endotracheal tube greatly contributes to Raw. The
resistance of the tube is normally higher than that of the
remaining internal airways.
2062971-001
CARESCAPE Modular Monitors
333
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...