Messages
Message
Location
Explanation
What to do
●
BIS measurement
removed
●
al. area
The module or BISx is
disconnected.
●
Connect the module or BISx
to start the measurement.
●
BIS module error
●
al. area
The module is not working
properly.
●
Check cable and connections.
●
Replace the BISx.
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
BIS sensor check
failed
●
al. area
The sensor has not passed the
impedance check.
●
Check sensor placement and
attachment.
●
BIS sensor expired
●
al. area
The sensor date has expired, the
sensor has been used too many
times, or the validity time for the
sensor cannot be determined.
●
Replace the sensor.
●
Cable off
●
param.,
wavef.
Cable is disconnected from the
module.
●
Connect the cable.
●
Checking sensor
●
param.,
wavef.
Sensor check is in progress.
●
Wait until the check is over.
Check results are displayed.
●
DSC error
●
param.,
wavef.
The BISx is not communicating
or operating properly.
●
Test the BISx.
●
If the problem persists,
contact authorized service
personnel.
●
High BIS impedance
●
al. area
Electrode impedance is too high.
●
Check electrode connections.
●
Identical BIS modules
●
al. area
There are two or more BIS
modules in the system.
●
Remove all but one BIS
module.
●
Incompatible DSC
●
param.
The hardware/ software is not
compatible with the BISx.
●
Check and change the BISx.
●
Incompatible sensor
●
param.,
wavef.
The sensor used is not a BIS
sensor.
●
Make sure that you are using
a Covidien BIS sensor.
●
Module error
●
param.,
wavef.
The module is not working
properly.
●
Check cable and connections.
●
Replace the BISx.
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
No BIS sensor
●
No sensor
●
al. area
●
param.,
wavef.
BIS: The sensor is not connected
to the digital signal processing
unit BISx.
●
Check the connection
between sensor and BISx,
and sensor and PIC (patient
interface cable).
●
Poor signal
●
param.,
wavef.
The BIS cannot be calculated
because the SQI is below 50.
●
Check electrode connections.
502
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...