Messages
Message
Location
Explanation
What to do
●
No patients found
●
al. area
No patients were found when
searching the ADT server.
●
Verify or change search
criteria.
●
Manually enter demographic
information.
●
No printer selected
●
al. area
There is no printer selected on
the monitor.
●
Select a printer.
●
Patient admitted
●
al. area
ICU, NICU and ED software
packages: The current patient
has just been admitted.
●
No action required.
●
Patient discharged
●
al. area
ICU, NICU and ED software
packages: The patient has just
been discharged.
●
No action required.
●
PDM battery low
●
al. area
PDM: The PDM battery cannot
be charged due to a power fault.
●
Allow PDM battery to charge.
●
If message persists, change
battery.
●
If the problem persists,
contact authorized service
personnel.
●
PDM battery temp
high
●
al. area
PDM: PDM battery temperature
is over the limit.
●
Contact authorized service
personnel.
●
PDM charging is
denied
●
al. area
The PDM battery cannot be
charged because the internal
temperature of the monitor is
too high.
●
Contact authorized service
personnel.
●
PDM module removed
●
al. area
Acquisition module has been
removed.
●
Connect the module if
you want to restart the
measurements.
●
PSM/PRESTN module
removed
●
al.area
Acquisition module has been
removed.
●
Connect the module if
you want to restart the
measurements.
●
Power management
failure
●
al. area
There is a problem with
communication to the power
management controller.
●
Wait and see if the message
disappears.
●
If the problem persists,
contact authorized service
personnel.
●
Printer error
●
al. area
Printer: A printer is not present
or the printer needs paper.
●
Select
Monitor Setup
>
Printing
to choose a different
printer.
●
Add paper to the printer.
●
Printing Alarm
●
al. area
Recorder: An alarm has
triggered printing.
●
Wait for the printing to finish.
●
Printing ready
●
al. area
Your printing request has been
forwarded to the printer.
●
No action required.
510
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...