Messages
Message
Location
Explanation
What to do
●
Wrong cable
●
param.
You have connected a
MasimoSet cable to the PDM
Nellcor, or a Nellcor OxiMax
cable to the PDM Masimo.
●
Check and use the correct
cable.
●
Wrong cable. Use
Masimo Set
●
al. area,
param.
You have connected a Nellcor
OxiMax cable to the PDM
Masimo.
●
Use a MasimoSet cable.
●
Wrong cable. Use
Nellcor OxiMax
●
al. area,
param.
You have connected a
MasimoSet cable to the PDM
Nellcor.
●
Use a Nellcor OxiMax cable.
Messages related to NIBP measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Air leakage
●
param.
Loose cuff or cuff hose.
●
Check the cuff and cuff hose.
●
Artifacts
●
param.
Measurement is disturbed by
artifact (e.g. patient movement,
shivering, deep breathing,
marked arrhythmia or irregular
beats).
●
Calm the patient and retry.
●
Calibrated
●
param.
Channel calibrated successfully.
●
Wait until the message
disappears before starting a
measurement.
●
Calibrating
●
param.
Calibration of a channel is in
progress.
●
No action required.
●
Calibration switch ON
●
param.
The calibration switch is on.
●
Contact authorized service
personnel.
●
Call service
●
param.
Technical fault.
●
Contact authorized service
personnel.
472
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...