Messages
Message
Location
Explanation
What to do
●
Agent mixture
●
al. area,
param.
Mixture of halogenated agents
is detected.
●
Check the ventilator and
agent vaporizer settings.
●
Apnea (CO2)
●
al. area,
wavef.
No breathing detected.
●
Check the patient status.
●
Check the ventilator and
breathing status.
●
Apnea deactivated
●
param.
A new case has just been
started/a new patient admitted,
or the measurement has just
been started and the apnea
alarm is not active yet.
●
Wait. The message
disappears after the monitor
detects 3 breaths during the
last minute (E-modules), or
the respiration rate is ≥3 per
minute (CAPNOSTAT, Dual
CO2, CapnoFlex).
●
Calibrate sensor
●
param.
A new sensor is introduced
or there is a signal drift in the
electronics.
●
Perform sensor calibration.
●
Calibrate system
●
param.
Interfaced device should be
calibrated.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Calibrating
●
param.
Calibration is in progress.
●
No action required.
●
Calibrating gas
sensor
●
wavef.
The measurement has been
started and calibration is in
progress.
●
Wait until the calibration is
completed and the message
disappears.
●
Calibration error
●
param.
Unsuccessful calibration.
●
Authorized service personnel
should repeat the calibration
procedure.
●
Check/calibr. adapter
●
al. area,
param.
The adapter is missing,
obstructed, or of a different
type than in the last calibration.
●
Check that there is an adapter
in the system.
●
Check the adapter for
blockages.
●
Calibrate the adapter.
●
Check ext device
●
param.
Interfaced device should be
checked.
●
Check the interfaced device.
●
If the problem persists,
contact authorized service
personnel.
●
Check sample gas out
●
al. area
The water trap is not connected,
the sample gas outflow is
blocked, or there is a leak inside
the module.
●
Check water trap connection.
●
Remove the blockage from
the sample gas outlet.
●
Change module if needed.
●
Check sample line
●
al. area,
param.
The sample line is disconnected,
or of a different type than the
last one used.
●
Reconnect the disconnected
sample line.
●
Calibrate the sample line.
2062971-001
CARESCAPE Modular Monitors
489
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...