Bispectral index
2.
Select the
Setup
tab.
3.
Select
Filters
. Set the filter option according to your needs: set the filters on by
selecting the check box, or set them off by deselecting the check box.
Setting BIS alarm limits
You can set the alarm limits according to your needs.
1.
Press the
module key or select the BIS parameter window.
2.
Select the
Alarms
tab.
NOTE
If the feature is not active, the alarm limits are greyed out.
Select
Alarm On
to set the alarms.
3.
Set the high and low alarm limits with the arrows.
Using the automatic BIS sensor check
1.
Press the
module key or select the BIS parameter window.
2.
Select the
Setup
tab.
3.
Select
Sensor Check
>
Automatic
.
Using the manual BIS sensor check
Whenever required, you can perform the sensor check manually. Press the
module key, or:
1.
Select the BIS parameter window.
2.
Select the
Setup
tab.
3.
Select
Check Sensor
.
4.
Observe the results on the screen.
The measurement continues automatically after the sensor has passed the check.
Testing the BISx
In case the measurement does not work and checking the cables and sensors does
not help, make sure that the BISx functions properly.
1.
Press the
module key or select the BIS parameter window
2.
Select the
Setup
tab.
3.
Select
Test DSC
.
4.
Observe the results on the screen.
After the test is completed, the result shown is
PASS
or
FAIL
.
Stopping the BIS measurement
1.
Remove the BIS sensor from the patient.
2.
Disconnect the sensor from the patient interface cable.
3.
Discard the sensor.
398
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...