Messages
Message
Location
Explanation
What to do
●
Headbox off
●
param.
Headbox is not connected.
●
Connect the headbox.
●
High EMG
●
param.
High frontal muscle activity may
disturb the EEG measurement.
●
Calm the patient.
●
Identical EEG
modules
●
al. area
There are two or more EEG
modules in the system.
●
Remove all but one EEG
module.
●
Poor electr. contact
●
param.
Electrode impedance is over 5
kOhm.
●
Press the electrode to
improve connection.
●
Change the electrode.
●
Check the impedance.
Messages related to BIS measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Apply sensor
●
param.
The sensor connection to the
patient may be loose.
●
Press the BIS sensor
electrodes to improve
connection.
●
Artifact
●
param.,
wavef.
Signals contain noise or artifact.
●
Check sensor contact.
●
Remove sources of excessive
noise.
●
Automatic check off
●
param.
Automatic sensor check has
been turned off.
●
If required, activate the
automatic check.
●
BIS cable off
●
al. area
Cable is disconnected from the
module.
●
Connect the cable.
●
BIS DSC error
●
al. area
The BISx is not communicating
or operating properly.
●
Test the BISx.
●
If the problem persists,
contact authorized service
personnel.
●
BIS high
/
BIS low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Check the dosage of
anesthetics.
●
Adjust alarm limits if
necessary.
2062971-001
CARESCAPE Modular Monitors
501
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...