ECG
set of stored (learned) templates. Incremental template updating allows information
from each beat, that correlates over time, to be reflected in the associated template.
ST detection measurement limitations
●
ST values may be affected by such factors as some drugs or metabolic and
conduction disturbances.
●
Since ST is often calculated with a fixed delay from the J point, changes in heart
rate may affect ST.
●
The ST algorithm has been tested for accuracy of the ST segment data. The
significance of the ST segment changes needs to be determined by a physician.
ST detection points to note
●
TRAM: When the TRAM has stored ST data, and is then connected to the monitor,
a new reference QRS is saved to the monitor at the time of connection. The
reference QRS from the TRAM cannot be used because the reference QRS does
not have time stamp.
●
ST segment deviations are not displayed for patients with ventricular pacemakers
or if the rhythm is considered as from ventricular origin.
Starting the ST detection
1.
Select the ST parameter window or select the HR parameter window and the
ST
tab.
2.
Select
Setup
.
3.
Select
On
from the
ST Analysis
list.
Selecting leads to the ST window
1.
Select the ST parameter window.
2.
Select
Setup
.
3.
Select the leads for display from the
ST Window
list.
Choices are:
●
ST Leads
: Displays the first three ST leads. The ST lead with the greatest
deviation is also displayed in the parameter window to the right of the ST leads.
●
All Leads
: Displays anterior, inferior, and lateral lead groups.
●
Off
: No ST parameter window is displayed. Instead, two ECG waveforms are
displayed beside the HR field (if
ECG2
has been selected to the screen).
Changing the displayed ST leads
You can select the display order of the first, second, and third displayed ST lead.
1.
Select the ST parameter window.
2.
Select
Setup
.
3.
Select a lead from the
ST Leads
list.
Adjusting the ST point manually
This selection is not available with combination monitoring.
2062971-001
CARESCAPE Modular Monitors
183
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...