ECG
time, arrhythmia detection may not be available. If the monitor is not able to relearn
due to a low amplitude QRS, for example, the
Arrhythmia paused
alarm is triggered.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select
Relearn QRS
.
Automatic relearning takes place when:
●
The measurement mode changes between the 3–lead mode and any other lead
mode.
●
The
ECG 1 Lead
selection is changed in the 3–lead mode.
●
The Va lead selection is changed in the 5– and 6–lead modes.
●
The ECG cable is connected (PSM, PDM).
●
The
Lead Analysis
setting is changed from
Multi lead
to
Single lead
.
Setting the primary HR source
The primary heart rate can be calculated from the ECG leads, SpO
2
measurement,
or invasive pressure waveform.
NOTE
This setting adjusts the primary heart rate source for all of the
hemodynamic parameters.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select a parameter from the
Primary HR Source
list. The selection list will only
show active measurements and
AUTO
or
IntelliRate
. Choices are (
UAC
with the
NICU software package only):
●
PDM with single HR:
IntelliRate
,
ECG
,
Art
,
ABP
,
Fem
,
UAC
,
Pleth
.
●
PDM with multiple HR:
IntelliRate
,
ECG
.
●
E-modules, TRAM, telemetry transmitters with single HR:
AUTO
,
ECG
,
Art
,
ABP
,
Fem
,
UAC
,
Pleth
.
●
E-modules, TRAM, telemetry transmitters with multiple HR:
AUTO
,
ECG
.
Showing a second HR value in the HR parameter window
You can display a second heart rate source in the HR parameter window.
1.
Select the HR parameter window.
2.
Select the
Advanced
tab.
3.
Select the
Show 2nd HR Source
check box to display the second HR source.
●
If the primary HR source for E-modules is
ECG
or
AUTO
(ECG), the secondary HR
source is displayed in this order:
Art
,
ABP
,
Fem
,
SpO2
.
●
If the primary HR source for PDM is
ECG
or
IntelliRate
(ECG), or
AUTO
(ECG) for
TRAM, the secondary HR source displayed in this order:
UAC
,
Art
,
ABP
,
Fem
,
SpO2
.
UAC
is available in the NICU software package only.
●
If the primary HR source is anything else than mentioned above, the secondary HR
source is always
ECG
.
162
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...