ECG
1.
Select the HR parameter window.
2.
Select the heart rate alarm range from the
HR Alarm Range
list:
●
30–240
(set and disabled if the
Primary HR Source
is
IntelliRate
,
AUTO
, or
SpO2
).
●
20–300
NOTE
France only: Available selections are
30–230
and
20–230
.
Selection
30–230
is set and disabled if the
Primary HR
Source
is
IntelliRate
,
AUTO
, or
SpO2
.
ECG alarm priorities
You can set the alarm priorities for various ECG alarms through
Alarm Setup
>
Alarm
Priorities
>
ECG
. The allowed priorities are defined in the
Care Unit Settings
, and
they are password protected.
ECG measurement practicalities
Alternate pulse rate source
The alternate pulse rate source allows clinicians to acquire a pulse rate from a source
other than ECG (Art, Fem, ABP, UAC, or SpO
2
). The following circumstances may
warrant the use of an alternate pulse rate source:
●
Excessive artifact due to an electrical interference from equipment (e.g.,
electrosurgical device).
●
Excessive patient movement causing significant artifact (e.g., seizure activity).
●
Inability to use standard lead placement (e.g., burns).
PDM IntelliRate algorithm
The PDM uses the
IntelliRate
algorithm.
IntelliRate
extracts information from multiple
physiological signals (ECG, SpO
2
, Art) and applies rule-based logic to determine which
heart rate source has the highest likelihood of being accurate. By reporting the most
accurate rate, the trended pulse rate is more accurate, and occurrences of false pulse
rate limit violation alarms are greatly reduced. The alternate pulse rate source value
replaces the standard heart rate value in the HR parameter window.
TRAM, E-modules, and telemetry transmitters Auto
algorithm
TRAM, E-modules, and telemetry transmitters use the
AUTO
algorithm.
AUTO
selects
the first available heart rate source based on a pre-defined parameter priority:
1.
ECG
2.
UAC (TRAM and NICU software package only)
3.
Art
4.
ABP
5.
Fem
6.
SpO
2
2062971-001
CARESCAPE Modular Monitors
165
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...