Messages
Message
Location
Explanation
What to do
●
Replace the module.
●
If the problem persists,
contact authorized service
personnel.
●
ABP disconnect
●
al. area
Invasive pressure line is
disconnected.
●
Check the patient status.
●
Check connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
ABP sys high
/
ABP sys
low
●
ABP mean high
/
ABP mean low
●
ABP dia high
/
ABP dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Art disconnect
●
al. area
Invasive pressure line is
disconnected.
●
Check the patient status.
●
Check connections.
●
If pressure drops because of
zeroing, perform the zeroing
process.
●
Art sys high
/
Art sys
low
●
Art mean high
/
Art mean low
●
Art dia high
/
Art dia
low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
●
Artifact
●
param.
If Smart BP is enabled, this
is normal behavior when
zeroing, flushing, or sampling is
performed.
If Smart BP is not enabled, this
message indicates that the
measurement is disturbed by
artifact.
●
Check the patient status.
●
Check cable contact.
●
Minimize tubing length.
●
Rezero the transducer.
●
Calibrated
●
param.
Channel calibrated successfully.
●
Wait until the message
disappears before starting a
measurement.
●
Calibrating
●
param.
Calibration of a channel is in
progress.
●
No action required.
●
Calibration failed
●
param.
Unsuccessful calibration.
●
Check the connections and
recalibrate.
●
CPP high
/
CPP low
●
al. area
Measurement values are equal
to or outside the alarm limits.
●
Check the patient status.
●
Adjust alarm limits if
necessary.
476
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...