Invasive pressures
Displayed values will switch between: Art 123/51 ( ) and Art 100/60 ( )
Invasive pressure troubleshooting
Problem
Solution
Artifact detected and the
Smart BP
option is turned
on.
●
Check the patient.
●
Reposition the catheter.
●
Zero the transducer.
●
If problem persists, turn off the
Smart BP
option.
●
If the
Smart BP
option is turned off, use the audio
pause feature before drawing blood and before
zeroing to reduce nuisance alarms.
Invasive pressure readings seem unstable.
●
Make sure there are no air bubbles in the
transducer systems.
●
Flush and zero.
●
Place the transducer on the patient’s phlebostatic
axis.
---/--- (80) Systolic and diastolic pressure values do
not display.
This may be due to the
Smart BP
option detecting
artifact. When artifacts are detected, only
Mean
values are displayed.
●
Check the patient.
●
Turn off the
Smart BP
if required.
Invasive pressure waveform is displayed but no
numeric values are displayed.
●
Zero the channel. Invasive pressure numeric
values are displayed only for successfully zeroed
channels.
Zeroing of invasive pressure channel(s) fails.
●
Ensure that the channels are open to air.
The channel standby is not selectable, or it is
terminated without the user giving the activation
request.
If the pressures remain between 10 mmHg and
250 mmHg for 10 seconds or more, the selection is
disabled or the standby is terminated.
252
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...