Alarms
Technical alarms’ deactivation with the pause
audio key
Certain technical alarms can be deactivated with the pause audio key. The
Alarm
Deactivation
setting is configured in the
Care Unit Settings
and it is password
protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Technical alarms for which the deactivation with the pause audio key can be allowed
are:
●
ECG Leads Off
●
Art disconnect
●
ABP disconnect
●
Fem disconnect
●
UAC disconnect
●
SpO2 Probe Off
Apnea alarms’ deactivation with the pause audio
key
Apnea alarms can be deactivated with the pause audio key if the
Allow alarm
deactivation with the Audio Pause key for:
setting
Apnea (CO2/Imped)
is enabled in
the
Care Unit Settings
. This setting is password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Pause audio with combination monitoring
When using combination monitoring, the pause audio behavior is the following:
●
If the telemetry transmitter is in pause audio state, also the monitor will be in the
same state. You can cancel the pause audio at the bedside monitor by selecting
the pause audio key. This will not affect the telemetry device.
●
If the monitor’s own pause audio state ends before the telemetry transmitter’s
audio pause, the monitor will re-enter pause audio.
●
The pause audio started by the telemetry transmitter will end also at the monitor
when the transmitter’s pause audio ends.
Breakthrough alarms
The breakthrough alarms feature allows pre-defined and user-selectable alarms to
“break through” (interrupt) an
All Alarms Audio Off
or a 2 or 5 minute audible alarm
pause condition.
The
FiO2 low
,
EtO2 low
,
FiN2O high
, and
Ppeak high
alarms will always break through
when escalated to or activated at high priority alarm condition regardless of the
All
Alarms Audio Off
selection or any alarm pausing.
2062971-001
CARESCAPE Modular Monitors
139
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...