Temperature
Stopping the temperature measurement
1.
Select the temperature parameter window.
2.
Select
Measurement
>
Off
.
Temperature practicalities
●
Each temperature label can be changed to reflect the temperature measurement
site.
●
Individual temperature sites can be turned off.
●
Tblood
is obtained from a pulmonary artery catheter.
●
The difference between two temperature sites can be calculated and displayed.
●
The dual temperature cable allows a two-channel measurement. The temperature
cable switch can be set to 400 or 700 depending on the type of probe used. Since
only 400 is supported, make sure that you turn the switch to this position.
●
The signal input is a high-insulation port to ensure patient safety and to protect the
device during defibrillation and electrosurgery.
●
The monitor automatically calibrates the temperature measurements at startup:
every 10 minutes for PSM, and every minute for PDM and TRAM.
Temperature troubleshooting
Problem
Solution
Temperature measurement fails
●
Check that the probe adapter is properly
connected to the acquisition module.
●
Check that the probe is properly connected to the
probe adapter.
●
Check that you are using the correct probe for the
anatomical location being monitored
●
Use a probe that is compatible with your system.
●
Try using a known good probe in case the sensor
is damaged.
●
Check that the acquisition module is properly
connected to the monitor.
●
Check the patient connection.
●
Check that there are not two identical
measurement modules in the system.
●
Contact service if the problem continues.
2062971-001
CARESCAPE Modular Monitors
259
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...