Impedance respiration
Respiration measurement checks
1.
Check that the waveform and parameter value are displayed when the cable is
connected to the patient.
NOTE
There may also be a respiration rate value displayed in the
CO
2
parameter window. Only the value in the respiration
parameter window is measured from the impedance
respiration source.
Respiration measurement on the monitor screen
●
The spikes in the waveform indicate detected inspiration and expiration.
●
PDM and TRAM: A text similar to
APN 15 s
indicates the value to which the apnea
alarm delay is set. In this example, the value is set to 15 seconds. It means that the
apnea alarm will activate after 15 seconds from the last detected breath.
Using the respiration measurement
Turning on the respiration measurement
The respiration measurement does not start automatically, so you must select it on.
1.
Select the impedance respiration parameter window.
2.
Select the
Setup
tab.
3.
Select
Respiration Measurement
>
On
.
Selecting the respiration lead
NOTE
TRAM and PDM only. In addition, the
RL-LL
respiration lead is
available with PDM only.
1.
Select the impedance respiration parameter window.
2.
Select the
Setup
tab.
3.
Select lead
I
,
II
, or
RL-LL
from the selection list on the right. Lead selections are
presented as graphical icons.
Selecting the respiration waveform size manually
1.
Select the impedance respiration parameter window.
2.
Select the
Setup
tab.
3.
Select a value from the
Size
list.
The greater the value, the larger the waveform size.
Selecting the respiration waveform size automatically
NOTE
PDM and TRAM only.
You can automatically size the waveform to fit the available space.
1.
Select the impedance respiration parameter window.
196
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...