Calculations
data for hemodynamic calculations. However, C.O. or CCO and/or their indexed values
that are older than 15 minutes are not used as source data. Other input values (for
example, HR, PA Mean, CVP, Art Mean) used in the calculation are chosen from the
same time the sample is drawn or the cardiac output is measured. For oxygenation
calculations the monitor uses laboratory data as source data.
In oxygenation calculations, you can select any available arterial laboratory data
samples or any C.O. measurement (if no laboratory data but multiple C.O. values are
available) from the current patient case to be used as source data.
In ventilation calculations, you can select any arterial laboratory data samples from
the current patient case to be used as source data.
Selecting source data for oxygenation calculations
1.
Select
Data & Pages
.
2.
Select
Calculations
.
3.
Select the
Oxy
tab.
4.
Select
Edit Input
.
5.
Select the desired sample with corresponding time and date from the
Select Lab
Data
list.
Selecting source data for ventilation calculations
B850 and B650 only.
1.
Select
Data & Pages
.
2.
Select
Calculations
.
3.
Select the
Vent
tab.
4.
Select
Edit Input
.
5.
Select the desired sample with corresponding time and date from the
Select Lab
Data
list.
Estimated values in oxygenation calculations
In normal circumstances, about 3% of the total arterial oxygen content is dissolved in
the blood and 97% is hemoglobin bound. When no SaO
2
laboratory result is saved
in the
Laboratory Data
menu, the measured SpO
2
value is used to estimate the
clinically relevant SaO
2
value. Also, the measured EtCO
2
value is used to estimate the
PaCO
2
value.
The monitor marks the estimated values by adding the letter e to the SaO
2
and PaCO
2
values in the
Calculations
>
Oxy
>
Trend
and
Calculations
>
Oxy
>
View
.
Estimated values in hemodynamic calculations
You can select different sources for the PCWP. If you select
LAP mean
or
PA
, the
PCWP
value in
Hemo
>
Trend
will be marked as an estimated value with the letter e. Instead,
the
Hemo
>
Edit Input
and
Hemo
>
View
show the actual selected label.
406
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...