Messages
Message
Location
Explanation
What to do
●
Zeroing
●
param.
IP channel is currently being
zeroed.
●
No action required.
Message is automatically
removed and replaced with the
zeroing results after completion.
●
Zero ICP separately
●
al. area
The ICP channel must be
zeroed separately from all other
invasive pressures.
●
Zero the channel using the
Zero
option found under the
ICP channel setup menu.
Messages related to temperature measurement
For information regarding alarm priorities and escalation times, see the CARESCAPE Modular
Monitors Supplemental Information Manual.
Make sure you are familiar with the generic layout of the screen. This will help you identify
where on screen the following messages appear. The message location is indicated with the
following abbreviations:
●
al. area = alarm area
●
param. = parameter window
●
report = report view
●
wavef. = waveform area
Message
Location
Explanation
What to do
●
Calibration check
●
param.
There is a 0.1°C difference
between the measured
temperature and the internal
calibration for the indicated
channel.
●
Change the cable.
●
Change the module.
●
If the problem persists,
contact authorized service
personnel.
●
Calibration fail
●
param.
Calibration on the indicated
channel failed.
●
Check connections.
●
Replace the transducer.
●
If the problem persists,
contact authorized service
personnel.
●
Identical temperature
modules
●
al. area
There are two or more identical
temperature modules in the
system.
●
Remove all but one
temperature module.
●
No sensor detected
●
param.
No sensor detected.
●
Check the sensor and
connections.
●
Performing temp test
●
param.
Module is calibrating.
●
No action required.
●
T1 Calibration check
/
T2 Calibration check
●
al. area
There is a 0.1°C difference
between the measured
temperature and the internal
calibration for the indicated
channel.
●
Change the cable.
●
Change the module.
●
If the problem persists,
contact authorized service
personnel.
2062971-001
CARESCAPE Modular Monitors
481
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...