DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-26
NOTE
Alarm silence timer is to mute the alarm according to the
timer setting by user. Alarm silence timer setting is accessible
in Misc. 1 screen.
NOTE
When overtemp alarm occurs, check the thermostat error
indicator light. Press on the red button on the thermostat to
reset it. Refer section
4.10 Recover from Over Temperature
more detail.
NOTE
If Cycle Complete Buzzer Alert Option is enabled, buzzer will
trigger after the cycle is complete.
This feature can be enabled/disabled in Misc. 1 screen.
NOTE
During cleaning process when
overtemp alarm
,
liquid leak
alarm
or
liquid low level
alarm is triggered, auto cycle of the
affected station will be aborted automatically. Alarm has to be
reset before pressing
[CYCLE START]
button to start Auto
cycle again.
NOTE
“
*OPTIONAL
” wording will be displayed on the touch
screen when the particular option is not purchased by user or
the option is not enabled in the Setup 6 screen.
Note
: User can enable/disable options if applicable in the
Setup 6 screen by access level as Administrator.
NOTE
When
Auto
mode process is running, user is not allowed to
change recipe. To change to other recipe, wait until current
process ends, alter the recipe number at Auto Screen, then
press
[CYCLE START]
button to begin a new auto cycle.
NOTE
Alarm will prompt when liquid leak is detected during
cleaning process.