DMS-R-IM
Operation
April 7, 2021
Digital Modular System - Rinse
4-27
NOTE
During Auto mode, heater and pump will turn ON
automatically if they are enabled in Auto Options screen. At
the same time heater and pump indicator in Manual Function
screen will turn ON too.
NOTE
During Manual mode, operator is unable to press
[CYCLE
START]
button to start auto cycle. Operator need to change to
Auto mode in Auto Screen.
NOTE
Make sure front and rear panels are closed properly and
[EMO]
button is released before press [SYSTEM ON] button
at touch screen.
Follow instruction below to begin Rinse Station cleaning cycle.
Step
Description
Images illustration
1
Fill tank with DI water/chemical
and ensure fill and drain ports are
connected to facility.
N/A
2
Set the process parameters
accordingly as mentioned in
section
4.8.1 Rinse Station Process Parameters
Setup.
N/A
3
Go to touch screen, at Main Menu
screen, login as Engineer level.
Refer section
for
procedure
4
Press [Auto Screen] button to enter
Auto Screen.