DMS-R-IM
Safety
April 7, 2021
Digital Modular System - Rinse
1-17
1.12.1
Panel Door Switch Interlock
Figure 1-3: System Rear View –Safety Panel Switch in electrical enclosure of each station
Panel Door Switches are installed in front control panel and electrical enclosure
panel on the rear side. It is used as the equipment’s safety interlocking device.
The electricity to the system is shut off when one of these switches is triggered
(when the panel is open). The system shut down similarly to when EMO button
is triggered (refer to section
1.11 Emergency OFF (EMO) Circuit
When station electrical panel is opened, the equipment will:
Stop all cleaning process. Pumps, heaters and ultrasonic generators will
be de-energized.
Alarm buzzer will turn ON.
To reset the interlock alarm:
1.
Close rear panel.
2.
Release EMO button.
3.
Press [SYSTEM ON] button on the touch screen to turn on the system.
Panel Door
Switch