DMS-R-IM
Interpreting HMI
April 7, 2021
Digital Modular System - Rinse
5-22
No.
Description
*
3
Manual Timed Drain Weir Setting (sec)
- This is drain weir time
setting to drain the remaining water in weir during
Manual
Mode
operation. It only available when weir pump option is selected. Weir
drain solenoid valve turns on when [DRAIN WEIR] button is pressed
in Manual Function screen. The drain solenoid valve will turn off
automatically when this drain time is complete.
Note
: Maximum Manual Timed Drain Weir Setting settable limit is 180
secs.
Note
: “
*
” is optional item.
5.7.5
Misc. 5 Screen
Figure 5-15: Misc. 5 screen (DMS-1218 & DMS-1826)
No.
Description
*
1
pH (pH)
Real time displays pH reading (pH).
Low Alarm Limit Set Point (pH)
– Set low pH alarm set point.
Alarm message “LOW pH LEVEL ALARM” will prompt when pH
reading is lower than this set point.
High Alarm Limit Set Point (pH)
– Set high pH alarm set point.
Alarm message “HIGH pH LEVEL ALARM” will prompt when pH
reading is higher than this set point.
Note
: Setting range of pH low and high limit set point is 0 – 14 pH.
1
2
3